Journal of Clinical and Diagnostic Research, ISSN - 0973 - 709X

Users Online : 12184

AbstractMaterial and MethodsResultsDiscussionConclusionKey MessageAcknowledgementReferences
Article in PDF How to Cite Citation Manager Readers' Comments (0) Audio Visual Article Statistics Link to PUBMED Print this Article Send to a Friend
Advertisers Access Statistics Resources

Dr Mohan Z Mani

"Thank you very much for having published my article in record time.I would like to compliment you and your entire staff for your promptness, courtesy, and willingness to be customer friendly, which is quite unusual.I was given your reference by a colleague in pathology,and was able to directly phone your editorial office for clarifications.I would particularly like to thank the publication managers and the Assistant Editor who were following up my article. I would also like to thank you for adjusting the money I paid initially into payment for my modified article,and refunding the balance.
I wish all success to your journal and look forward to sending you any suitable similar article in future"



Dr Mohan Z Mani,
Professor & Head,
Department of Dermatolgy,
Believers Church Medical College,
Thiruvalla, Kerala
On Sep 2018




Prof. Somashekhar Nimbalkar

"Over the last few years, we have published our research regularly in Journal of Clinical and Diagnostic Research. Having published in more than 20 high impact journals over the last five years including several high impact ones and reviewing articles for even more journals across my fields of interest, we value our published work in JCDR for their high standards in publishing scientific articles. The ease of submission, the rapid reviews in under a month, the high quality of their reviewers and keen attention to the final process of proofs and publication, ensure that there are no mistakes in the final article. We have been asked clarifications on several occasions and have been happy to provide them and it exemplifies the commitment to quality of the team at JCDR."



Prof. Somashekhar Nimbalkar
Head, Department of Pediatrics, Pramukhswami Medical College, Karamsad
Chairman, Research Group, Charutar Arogya Mandal, Karamsad
National Joint Coordinator - Advanced IAP NNF NRP Program
Ex-Member, Governing Body, National Neonatology Forum, New Delhi
Ex-President - National Neonatology Forum Gujarat State Chapter
Department of Pediatrics, Pramukhswami Medical College, Karamsad, Anand, Gujarat.
On Sep 2018




Dr. Kalyani R

"Journal of Clinical and Diagnostic Research is at present a well-known Indian originated scientific journal which started with a humble beginning. I have been associated with this journal since many years. I appreciate the Editor, Dr. Hemant Jain, for his constant effort in bringing up this journal to the present status right from the scratch. The journal is multidisciplinary. It encourages in publishing the scientific articles from postgraduates and also the beginners who start their career. At the same time the journal also caters for the high quality articles from specialty and super-specialty researchers. Hence it provides a platform for the scientist and researchers to publish. The other aspect of it is, the readers get the information regarding the most recent developments in science which can be used for teaching, research, treating patients and to some extent take preventive measures against certain diseases. The journal is contributing immensely to the society at national and international level."



Dr Kalyani R
Professor and Head
Department of Pathology
Sri Devaraj Urs Medical College
Sri Devaraj Urs Academy of Higher Education and Research , Kolar, Karnataka
On Sep 2018




Dr. Saumya Navit

"As a peer-reviewed journal, the Journal of Clinical and Diagnostic Research provides an opportunity to researchers, scientists and budding professionals to explore the developments in the field of medicine and dentistry and their varied specialities, thus extending our view on biological diversities of living species in relation to medicine.
‘Knowledge is treasure of a wise man.’ The free access of this journal provides an immense scope of learning for the both the old and the young in field of medicine and dentistry as well. The multidisciplinary nature of the journal makes it a better platform to absorb all that is being researched and developed. The publication process is systematic and professional. Online submission, publication and peer reviewing makes it a user-friendly journal.
As an experienced dentist and an academician, I proudly recommend this journal to the dental fraternity as a good quality open access platform for rapid communication of their cutting-edge research progress and discovery.
I wish JCDR a great success and I hope that journal will soar higher with the passing time."



Dr Saumya Navit
Professor and Head
Department of Pediatric Dentistry
Saraswati Dental College
Lucknow
On Sep 2018




Dr. Arunava Biswas

"My sincere attachment with JCDR as an author as well as reviewer is a learning experience . Their systematic approach in publication of article in various categories is really praiseworthy.
Their prompt and timely response to review's query and the manner in which they have set the reviewing process helps in extracting the best possible scientific writings for publication.
It's a honour and pride to be a part of the JCDR team. My very best wishes to JCDR and hope it will sparkle up above the sky as a high indexed journal in near future."



Dr. Arunava Biswas
MD, DM (Clinical Pharmacology)
Assistant Professor
Department of Pharmacology
Calcutta National Medical College & Hospital , Kolkata




Dr. C.S. Ramesh Babu
" Journal of Clinical and Diagnostic Research (JCDR) is a multi-specialty medical and dental journal publishing high quality research articles in almost all branches of medicine. The quality of printing of figures and tables is excellent and comparable to any International journal. An added advantage is nominal publication charges and monthly issue of the journal and more chances of an article being accepted for publication. Moreover being a multi-specialty journal an article concerning a particular specialty has a wider reach of readers of other related specialties also. As an author and reviewer for several years I find this Journal most suitable and highly recommend this Journal."
Best regards,
C.S. Ramesh Babu,
Associate Professor of Anatomy,
Muzaffarnagar Medical College,
Muzaffarnagar.
On Aug 2018




Dr. Arundhathi. S
"Journal of Clinical and Diagnostic Research (JCDR) is a reputed peer reviewed journal and is constantly involved in publishing high quality research articles related to medicine. Its been a great pleasure to be associated with this esteemed journal as a reviewer and as an author for a couple of years. The editorial board consists of many dedicated and reputed experts as its members and they are doing an appreciable work in guiding budding researchers. JCDR is doing a commendable job in scientific research by promoting excellent quality research & review articles and case reports & series. The reviewers provide appropriate suggestions that improve the quality of articles. I strongly recommend my fraternity to encourage JCDR by contributing their valuable research work in this widely accepted, user friendly journal. I hope my collaboration with JCDR will continue for a long time".



Dr. Arundhathi. S
MBBS, MD (Pathology),
Sanjay Gandhi institute of trauma and orthopedics,
Bengaluru.
On Aug 2018




Dr. Mamta Gupta,
"It gives me great pleasure to be associated with JCDR, since last 2-3 years. Since then I have authored, co-authored and reviewed about 25 articles in JCDR. I thank JCDR for giving me an opportunity to improve my own skills as an author and a reviewer.
It 's a multispecialty journal, publishing high quality articles. It gives a platform to the authors to publish their research work which can be available for everyone across the globe to read. The best thing about JCDR is that the full articles of all medical specialties are available as pdf/html for reading free of cost or without institutional subscription, which is not there for other journals. For those who have problem in writing manuscript or do statistical work, JCDR comes for their rescue.
The journal has a monthly publication and the articles are published quite fast. In time compared to other journals. The on-line first publication is also a great advantage and facility to review one's own articles before going to print. The response to any query and permission if required, is quite fast; this is quite commendable. I have a very good experience about seeking quick permission for quoting a photograph (Fig.) from a JCDR article for my chapter authored in an E book. I never thought it would be so easy. No hassles.
Reviewing articles is no less a pain staking process and requires in depth perception, knowledge about the topic for review. It requires time and concentration, yet I enjoy doing it. The JCDR website especially for the reviewers is quite user friendly. My suggestions for improving the journal is, more strict review process, so that only high quality articles are published. I find a a good number of articles in Obst. Gynae, hence, a new journal for this specialty titled JCDR-OG can be started. May be a bimonthly or quarterly publication to begin with. Only selected articles should find a place in it.
An yearly reward for the best article authored can also incentivize the authors. Though the process of finding the best article will be not be very easy. I do not know how reviewing process can be improved. If an article is being reviewed by two reviewers, then opinion of one can be communicated to the other or the final opinion of the editor can be communicated to the reviewer if requested for. This will help one’s reviewing skills.
My best wishes to Dr. Hemant Jain and all the editorial staff of JCDR for their untiring efforts to bring out this journal. I strongly recommend medical fraternity to publish their valuable research work in this esteemed journal, JCDR".



Dr. Mamta Gupta
Consultant
(Ex HOD Obs &Gynae, Hindu Rao Hospital and associated NDMC Medical College, Delhi)
Aug 2018




Dr. Rajendra Kumar Ghritlaharey

"I wish to thank Dr. Hemant Jain, Editor-in-Chief Journal of Clinical and Diagnostic Research (JCDR), for asking me to write up few words.
Writing is the representation of language in a textual medium i e; into the words and sentences on paper. Quality medical manuscript writing in particular, demands not only a high-quality research, but also requires accurate and concise communication of findings and conclusions, with adherence to particular journal guidelines. In medical field whether working in teaching, private, or in corporate institution, everyone wants to excel in his / her own field and get recognised by making manuscripts publication.


Authors are the souls of any journal, and deserve much respect. To publish a journal manuscripts are needed from authors. Authors have a great responsibility for producing facts of their work in terms of number and results truthfully and an individual honesty is expected from authors in this regards. Both ways its true "No authors-No manuscripts-No journals" and "No journals–No manuscripts–No authors". Reviewing a manuscript is also a very responsible and important task of any peer-reviewed journal and to be taken seriously. It needs knowledge on the subject, sincerity, honesty and determination. Although the process of reviewing a manuscript is a time consuming task butit is expected to give one's best remarks within the time frame of the journal.
Salient features of the JCDR: It is a biomedical, multidisciplinary (including all medical and dental specialities), e-journal, with wide scope and extensive author support. At the same time, a free text of manuscript is available in HTML and PDF format. There is fast growing authorship and readership with JCDR as this can be judged by the number of articles published in it i e; in Feb 2007 of its first issue, it contained 5 articles only, and now in its recent volume published in April 2011, it contained 67 manuscripts. This e-journal is fulfilling the commitments and objectives sincerely, (as stated by Editor-in-chief in his preface to first edition) i e; to encourage physicians through the internet, especially from the developing countries who witness a spectrum of disease and acquire a wealth of knowledge to publish their experiences to benefit the medical community in patients care. I also feel that many of us have work of substance, newer ideas, adequate clinical materials but poor in medical writing and hesitation to submit the work and need help. JCDR provides authors help in this regards.
Timely publication of journal: Publication of manuscripts and bringing out the issue in time is one of the positive aspects of JCDR and is possible with strong support team in terms of peer reviewers, proof reading, language check, computer operators, etc. This is one of the great reasons for authors to submit their work with JCDR. Another best part of JCDR is "Online first Publications" facilities available for the authors. This facility not only provides the prompt publications of the manuscripts but at the same time also early availability of the manuscripts for the readers.
Indexation and online availability: Indexation transforms the journal in some sense from its local ownership to the worldwide professional community and to the public.JCDR is indexed with Embase & EMbiology, Google Scholar, Index Copernicus, Chemical Abstracts Service, Journal seek Database, Indian Science Abstracts, to name few of them. Manuscriptspublished in JCDR are available on major search engines ie; google, yahoo, msn.
In the era of fast growing newer technologies, and in computer and internet friendly environment the manuscripts preparation, submission, review, revision, etc and all can be done and checked with a click from all corer of the world, at any time. Of course there is always a scope for improvement in every field and none is perfect. To progress, one needs to identify the areas of one's weakness and to strengthen them.
It is well said that "happy beginning is half done" and it fits perfectly with JCDR. It has grown considerably and I feel it has already grown up from its infancy to adolescence, achieving the status of standard online e-journal form Indian continent since its inception in Feb 2007. This had been made possible due to the efforts and the hard work put in it. The way the JCDR is improving with every new volume, with good quality original manuscripts, makes it a quality journal for readers. I must thank and congratulate Dr Hemant Jain, Editor-in-Chief JCDR and his team for their sincere efforts, dedication, and determination for making JCDR a fast growing journal.
Every one of us: authors, reviewers, editors, and publisher are responsible for enhancing the stature of the journal. I wish for a great success for JCDR."



Thanking you
With sincere regards
Dr. Rajendra Kumar Ghritlaharey, M.S., M. Ch., FAIS
Associate Professor,
Department of Paediatric Surgery, Gandhi Medical College & Associated
Kamla Nehru & Hamidia Hospitals Bhopal, Madhya Pradesh 462 001 (India)
E-mail: drrajendrak1@rediffmail.com
On May 11,2011




Dr. Shankar P.R.

"On looking back through my Gmail archives after being requested by the journal to write a short editorial about my experiences of publishing with the Journal of Clinical and Diagnostic Research (JCDR), I came across an e-mail from Dr. Hemant Jain, Editor, in March 2007, which introduced the new electronic journal. The main features of the journal which were outlined in the e-mail were extensive author support, cash rewards, the peer review process, and other salient features of the journal.
Over a span of over four years, we (I and my colleagues) have published around 25 articles in the journal. In this editorial, I plan to briefly discuss my experiences of publishing with JCDR and the strengths of the journal and to finally address the areas for improvement.
My experiences of publishing with JCDR: Overall, my experiences of publishing withJCDR have been positive. The best point about the journal is that it responds to queries from the author. This may seem to be simple and not too much to ask for, but unfortunately, many journals in the subcontinent and from many developing countries do not respond or they respond with a long delay to the queries from the authors 1. The reasons could be many, including lack of optimal secretarial and other support. Another problem with many journals is the slowness of the review process. Editorial processing and peer review can take anywhere between a year to two years with some journals. Also, some journals do not keep the contributors informed about the progress of the review process. Due to the long review process, the articles can lose their relevance and topicality. A major benefit with JCDR is the timeliness and promptness of its response. In Dr Jain's e-mail which was sent to me in 2007, before the introduction of the Pre-publishing system, he had stated that he had received my submission and that he would get back to me within seven days and he did!
Most of the manuscripts are published within 3 to 4 months of their submission if they are found to be suitable after the review process. JCDR is published bimonthly and the accepted articles were usually published in the next issue. Recently, due to the increased volume of the submissions, the review process has become slower and it ?? Section can take from 4 to 6 months for the articles to be reviewed. The journal has an extensive author support system and it has recently introduced a paid expedited review process. The journal also mentions the average time for processing the manuscript under different submission systems - regular submission and expedited review.
Strengths of the journal: The journal has an online first facility in which the accepted manuscripts may be published on the website before being included in a regular issue of the journal. This cuts down the time between their acceptance and the publication. The journal is indexed in many databases, though not in PubMed. The editorial board should now take steps to index the journal in PubMed. The journal has a system of notifying readers through e-mail when a new issue is released. Also, the articles are available in both the HTML and the PDF formats. I especially like the new and colorful page format of the journal. Also, the access statistics of the articles are available. The prepublication and the manuscript tracking system are also helpful for the authors.
Areas for improvement: In certain cases, I felt that the peer review process of the manuscripts was not up to international standards and that it should be strengthened. Also, the number of manuscripts in an issue is high and it may be difficult for readers to go through all of them. The journal can consider tightening of the peer review process and increasing the quality standards for the acceptance of the manuscripts. I faced occasional problems with the online manuscript submission (Pre-publishing) system, which have to be addressed.
Overall, the publishing process with JCDR has been smooth, quick and relatively hassle free and I can recommend other authors to consider the journal as an outlet for their work."



Dr. P. Ravi Shankar
KIST Medical College, P.O. Box 14142, Kathmandu, Nepal.
E-mail: ravi.dr.shankar@gmail.com
On April 2011
Anuradha

Dear team JCDR, I would like to thank you for the very professional and polite service provided by everyone at JCDR. While i have been in the field of writing and editing for sometime, this has been my first attempt in publishing a scientific paper.Thank you for hand-holding me through the process.


Dr. Anuradha
E-mail: anuradha2nittur@gmail.com
On Jan 2020

Important Notice

Original article / research
Year : 2010 | Month : April | Volume : 4 | Issue : 2 | Page : 2163 - 2168 Full Version

Reliability of Capnometry in Neonates on Patient Triggered Ventilation


Published: April 1, 2010 | DOI: https://doi.org/10.7860/JCDR/2010/.680
PARIKH T B * , NANAVATI RN**, UDANI R H***

*(DM), ** (MD), ***(MD), Dept of Neonatology, Seth GS Medical College & King Edward Memorial Hospital Parel, Mumbai, (India)

Correspondence Address :
Dr.Tushar B. Parikh (DNB),(DM)(Neonatology)Consultant & Incharge NICU,Oyster and Pearl Hospital, Pune
Mailing address:5/5, Five star co-op Society,Near Hotel Sun-n-sand Bundgarden Road, Pune -411001,E mail – drtusharparikh@gmail.com

Abstract

Introduction: Capnometry is not well studied in neonates on Patient Triggered Ventilation (PTV). We conducted this study to determine the reliability of Main Stream (MS) and Side Stream (SS) Capnometry in neonates on PTV.
Method: Neonates on PTV were enrolled in the study. Before each Arterial Blood Gas (ABG) assessment, MS and SS capnometer readings were recorded. Lung mechanics were recorded from the online graphics. ABG was collected from the arterial line. The EtCO2 – PaCO2 correlation was drawn for MS and SS capnometry under various disease conditions, ventilator settings and lung mechanics.
Results: A total of 74 ABGs were collected from 18 patients. The EtCO2 – PaCO2 correlation was better with MS capnometry than with SS capnometry (r = 0.855 vs.0.68, p < 0.001 for both methods). Both methods correlated well with PaCO2 in Flow cycled SIMV (MS: r = 0.9 & SS: r = 0.82). However, in Flow Cycled Assist Control mode, SS capnometry correlated poorly (SS: r = 0.49 vs. MS: r = 0.76). The EtCO2 - PaCO2 correlation by MS capnometry holds good for RDS, Apnea, Pneumonia and Congenital cyanotic heart disease (r = 0.85, 0.97, 0.84, 0.84 respectively, p < 0.001), but not for PPHN(r = 0.37, p = 0.35). SS EtCO2 correlated well in RDS, Apnea, Pneumonia, (r = 0.75, 0.85, 0.94 respectively, p < 0.001), but not in PPHN (- 0.20, p = 0.629) and CCHD (r = 0.73, p = 0.1). At higher ventilator rates (> 60 /min), SS capnometry correlated poorly. The EtCO2 –PaCO2 correlation by both methods was better when lung compliance was > 1 ml / cmH2O than when lung compliance was < 1 ml / cmH2O.
Conclusions: MS capnometry is superior to SS capnometry for neonates on patient triggered ventilation.

Keywords

End tidal capnometry (EtCO2), Patient triggered ventilation (PTV), lung mechanics

Introduction
End tidal capnometry (EtCO2) allows exhaled carbon-dioxide (CO2) to be measured non-invasively and continuously. EtCO2 is a measure of the alveolar partial pressure of CO2 (PACO2), which in turn is a measure of arterial PCO2 (PaCO2). Two types of capnometers which are in use are, Main stream (the CO2 sensors are located directly in the patient’s breathing circuit) and side stream (CO2 sensor located in the machine base unit) (1),(2). Till recently, capnometry was not embraced by neonatologists as a means to measure ventilation and lung mechanics, as the neonatal lung volumes are small and the dead space of the capnometer adapters / sensors was higher (3),(4),(5) Low dead-space MS CO2 analyzers for newborns have been available since the last decade and they have been found to be of clinical use (6),(7).

Patient Triggered Ventilation is characterized by complete respiratory synchrony between the patient and machine breaths, short inspiratory time (In flow cycled modes), better control over the delivered tidal volumes and ventilation at lower pressures. (8) SIMV allows spontaneous breaths between the machine breaths. PTV is rapidly replacing IMV as primary mode of ventilation in neonates. Because of lack of data, we evaluated the use of different capnometry methods in newborns on PTV.

Primary Objective
To determine the reliability of Main Stream (MS) and Side Stream (SS) capnometry in neonates on PTV

Secondary Outcome Measures:
1. To test the reliability of capnometry under various disease conditions, ventilator settings and lung mechanics.
2. Prediction of Pa CO2 by MS and SS capnometry.

Setting
Tertiary care Neonatal Intensive care unit

Inclusion Criteria
Newborns on patient triggered ventilation, having online pulmonary graphic display and an indwelling functioning arterial line were included.

Exclusion Criteria
Situations where urgent arterial blood gas reports are clinically indicated.

Methodology
All patients who were ventilated in the unit during three consecutive months were enrolled in the study. All babies were ventilated on the Bear 750 PSV ventilator on different patient triggered modes with flow trigger sensor (Hot wire anemometer) having capabilities of online continuous graphics monitoring of waveforms, loops and lung mechanics. Distal Main stream Et CO2 (MS Et CO2) monitoring was done by using the respiratory monitor CO2SMO+ (Novametrix Medical Systems,). The ‘flow through’ airway adapter had a dead space of 0.7 ml. Side stream Et CO2 (SS Et CO2) monitoring was done by inbuilt capnogram with the multichannel monitor, (Criticare Systems Inc.), with a sample aspiration rate of 50 ml/ min and an airway adapter with 0.6ml of dead space.

Material and Methods

ABG samples were collected as per the clinical condition of the baby and the unit policy. Before each investigation, in the MS capnograph, the CO2 sensor was calibrated with two independent calibration cells provided by the manufacturer. The SS capnometer was calibrated to zero as per the manufacturer’s instructions. Lung mechanics were recorded on the proforma. The SS capnograph reading was recorded for complete one minute and the mode value was taken as the “Side stream Et CO2†value. Then, the patient was disconnected from the side-port airway adapter and was connected to the MS EtCO2 adapter and the “Main stream EtCO2†value was recorded in a similar manner after a period of equilibration of 30 seconds or till a stable EtCO2 reading was displayed, whichever occurred early. Arterial blood gas (ABG) samples were collected immediately after EtCO2 measurement. ABG samples were transported on ice slush and were analyzed on an Instrumentation Laboratory Inc. (US) machine within 10 minutes of their collection. ABG, EtCO2 values and lung mechanics were recorded on a pre-designed proforma.

The data was entered in a computer and was statistically analyzed by using the SPSS ver.11 for windows. A statistical correlation between EtCO2 and PaCO2 was drawn for the MS and SS methods. The EtCO2-PaCO2 correlation was tested under different modes of PTV, various disease conditions (i.e., RDS, Apnea etc.), ventilator settings (viz. ventilator rate, mean airway pressure) and lung mechanics parameters (viz. lung compliance, tidal volume and minute ventilation.) The effect of categorical variables was assessed by drawing a correlation between EtCO2 and PaCO2 by paired t test, under different categories. The effect of continuous variables(eg. mean airway pressure) was assessed by calculating the EtCO2 – PaCO2 difference for the type of EtCO2 monitoring by paired t test. Ninety five percent confidence intervals were taken as significant.

Results

During the three month study period, 74 ABG samples with corresponding MS and SS capnographic readings and lung mechanics were recorded from 18 patients. The patients were ventilated for various lung conditions like, RDS, pneumonia, Apnea, Meconium aspiration with PPHN and congenital cyanotic heart disease. The gestational ages of the babies ranged from 27 weeks to 40 weeks and the birth weights ranged from 920 to 2600 grams.

I) Correlation between Et CO2 And Pa CO2 (Table/Fig 1)
The average SS Et CO2 reading [24.18 mmHg (SD = 6.06)] was significantly lower than the average MS Et CO2 reading [30.68 mmHg (SD = 6.24), p < 0.001].

The arterial-end-tidal PaCO2 difference for the MS method was 3.65 mmHg (CI +2.98 to +4.33), which was significantly less than the arterial – end tidal PCO2 difference for the SS method- 9.22 mmHg. (CI + 7.97 to + 10.47),p< 0.001.

II) Linear Regression Equation for Pa CO2 (Table/Fig 2)
MS Capnometry: PaCO2 = 4.01 + 0.96 (EtCO2 value)
SS Capnometry: PaCO2 = 24.94+ 0.35 (EtCO2 value)

III) Mode of Ventilation (Table/Fig 3)

IV) Correlation between Pa CO2 and Et CO2 For Various Disease Conditions (Table/Fig 4)

IV) Correlation between Arterial and End Tidal Pco2 under Various Ventilator Settings
1. Ventilator Breath Rate (Table/Fig 5)
2. The effect of the mean airway pressure on the EtCO2-PaCO2 correlation was assessed by the correlation between the mean airway pressure and the Arterial-end-tidal pCO2 difference. The mean reading for Mean airway pressure was 6.059 (+ 2.98). Changes in the mean airway pressure had a weak linear correlation with the Arterial-end-tidal pCO2 difference with the MS capnometry {3.65 (+ 5.82), r = 0.27, p = 0.017}; however, changes in the mean airway pressure had a moderately positive linear correlation with the Arterial-end-tidal pCO2 difference with SS capnometry, {9.22, (+ 10.78), r = 0.46, p < 0.001}, thus indicating that as the mean airway pressure increased, SS capnometry gave significantly lower readings.

V) Correlation with Lung Mechanics Parameters
1. Lung compliance: (Table/Fig 6)
2. Tidal Volume: (Table/Fig 7)

Discussion

This is the first attempt to systematically test the reliability of capnometers in neonates on PTV. This study shows that both the capnometry methods show good correlation with PaCO2; however, the MS method had better correlation coefficients. In MS capnometry, CO2 is estimated by the sensor placed in the breathing circuit, while in the SS method, the exhaled air is aspirated through the sample aspirator into the machine base unit. This could be the reason why CO2 changes were rapidly and more accurately reflected by the MS capnogram. In a previous study on the comparison of side-stream and mainstream capnometers in neonates by McEvedy et. al., the slope of the least square regression line for the distal side-stream capnometer- 0.67, was significantly less than that for the mainstream capnometer- 0.78 (10) Some of the other studies confirm these findings. (11), (12), (13)

During the Flow cycled SIMV mode of ventilation, both the methods of Capnometry gave good correlation with PaCO2 values (r= 0.8). However, under the Flow cycled Assist control mode, the correlation coefficient obtained by the SS method was much lower than that obtained by the MS method (0.76 vs. 0.497). Ventilation in the FCAC mode is characterized by the delivery of a set pressure limited flow cycled breath, with every breathing attempt, reaching the trigger sensitivity. There was no control over the maximum machine breath-rate delivery in this mode. The alveolar phase of expiration may not be complete when the breathing rates are high or when the patient exhales out of phase with the machine breath (12), (16) These rapidly changing volumes may not allow sufficient time for the aspiration of the representative sample in SS capnometry. This could be the possible explanation for poor correlation coefficients in SS capnometry under FCAC modes of ventilation. Similar data was not available for comparison.

EtCO2 monitoring was reliable with most of the disease conditions in neonates such as RDS, Apnea and Pneumonia. One previous study from India by Nangia et.al which was done on babies on conventional ventilation, showed a good correlation between EtCO2 and PaCO2, in babies with Recurrent Apnea and Meconium Aspiration syndrome (r = 0.96 & 0.94 respectively); however, the correlation coefficient was the lowest (0.55) for babies with RDS (15) In the present study, babies were ventilated on Patient Triggered modes of ventilation with a conscious effort to maintain the tidal volume between 5 to 8 ml/kg, which could have contributed to better correlation coefficients even in RDS.

EtCO2 did not correlate with the PaCO2 in Persistent Pulmonary Hypertension of the newborn (r = 0.37 & - 0.2, p =0.35 & 0.629, MS and SS respectively). In PPHN, due to the shunting of deoxygenated blood at the cardiac and extra-cardiac sites, there was ventilation perfusion mismatch; also, alkali therapy which is commonly used in PPHN may lead to the rapid rise in PaCO2-which may not correlate with EtCO2 (14),(15) The correlation coefficient was very low in patients with Congenital cyanotic heart disease. In a study by Lazzell VA et.al., the stability of the intra-operative arterial to end-tidal carbon dioxide partial pressure difference in children with congenital heart disease was assessed. It was seen that arterial end-tidal CO2 difference in children with acyanotic-shunting and cyanotic heart disease with intra-cardiac mixing (normal or increased pulmonary blood flow) was stable intra-operatively, although patients with cyanotic mixing congenital heart lesions (decreased or variable pulmonary blood flow) did demonstrate large individual variation. (14). Therefore, EtCO2 monitoring is not recommended in variable right to left shunt situations (PPHN) or in congenital cyanotic heart diseases with decreased or variable blood flow, as it is not very reliable (14)

Patients with higher ventilator rates (> 60 / min) had poor correlation of EtCO2 and PaCO2 by the SS method, (0.82 vs. 0.56). In a study by Pascucci RC et.al, side-stream and main-stream capnograms were compared in infants on rapid mechanical ventilation. The EtCO2-PaCO2 correlation was better with mainstream capnometers. The recordings obtained by the SS machine were grossly distorted, with flattening of the ascending limb (slope 37.3 vs.153.3 torr/sec, SS vs. MS, p less than 0.001) and absence of the alveolar plateau (17). The alveolar phase of expiration may not be complete when the ventilator rates are high (16), (17)

As the tidal volume increased (> 10 ml), the correlation by MS capnometry remained good (r = 0.84). However, the side port EtCO2 measurement decreased. (r = 0.59). Since, in case of SS capnometry, the exhaled air sample is aspirated in the machine base unit and the measurement of EtCO2 is done during extremes of tidal volumes, exhaled PCO2 may not equilibrate with the PCO2 in the aspirated sample. EtCO2 gave lower readings when the tidal volume was either too low or too high (17) .

As the lung compliance decreased below sub-physiological levels (< 1 ml/ cm H2O), the correlation of EtCO2 with PaCO2 decreased by both the methods of capnometry. This is because of the relative increased dead space ventilation with short time constants in diseases of decreased lung compliance (15).

Thus, When Used Properly, Etco2 Monitoring Along With Pulse-Ox Monitoring Has Great Potential To Reduce The Phlebotomy Losses In The Population Where It Matters Most!

Conclusion

Conclusions and Recommendations
(Key Messages)
1. MS capnometry is superior to SS capnometry in neonates on patient triggered ventilation, especially if ventilated in Flow cycled assist control mode or with rapid ventilator rate.
2. For practical purposes, predicted Pa CO2 by the MS method is PaCO2 = 4 + EtCO2 and by the SS method, it is PaCO2 = 25 +1 /3 EtCO2
3. Diseases with right to left shunt situation capnometry are less reliable.
4. We recommend using MS end tidal capnometry for neonates on the patient triggered mode of ventilation

Contributors
TBP was involved in concept, design, data collection and writing the primary draft. RNN was involved in design of the study and drafting the manuscript. RHU was involved in the overall supervision of the study.

Funding: None

Competing Interests: None

Key Message

Main stream capnometry is superior to side stream capnometry for neonates on patient triggered ventilation. In diseases with right to left shunt situation, capnometry may not be reliable. In ventilating diseases with poor lung compliance and high ventilator rates, capnometry correlates poorly to PaCO2.

Acknowledgement

Our sincere thanks to Dr.Chaitali Warang, Dr. Dhananjay Sangle and Dr.Rajeev Ranjan for their help in data collection, Dr.Naveen Bajaj for important suggestions during the conduction of study and Dr.Nandkishore Kabra for important suggestions in the manuscript. Acknowledgements are due to Dr. Nilima Kshirsagar, Dean, Seth GS Medical College & KEM Hospital, Mumbai for permitting to publish this manuscript.

References

1.
Bhende MS. End-tidal carbon dioxide monitoring in pediatrics: concepts and technology. J Postgrad Med 2001;47:153-6
2.
Gravenstein JHS, Paulus DA, Hayes TJ. Clinical indications. In: Gravenstein JS, Paulus DA, Hayes TJ (Eds) Capnometry in clinical practice. Stoneham, MA: Butterworth; 1989, pp 43-49
3.
Kirpalani H , Kechagias S , Lerman J : Technical and clinical aspects of capnography in neonates . J Med Eng Technol 1991 ; 15 : 154–61.
4.
Tirosh E, Bilker A, Bader D, Cohen A. Capnography in spontaneously breathing preterm and term infants . Clin Physiol 2001 ; 21 : 150–54
5.
Proquitte H, Krause S, Rudiger M, Wauer RR, Schmalisch G. Current limitations of volumetric capnography in surfactant-depleted small lungs. Pediatr Crit Care Med. 2004 Jan;5(1):75-80.
6.
Cannon ML, Cornell J, Tripp-Hamel DS, Gentile MA, Hubble CL, Meliones JN et.al. Tidal volumes for ventilated infants should be determined with a pneumotachometer placed at the endotracheal tube. Am J Respir Crit Care Med 2000; 162 : 2109–12
7.
Rozycki HJ, Sysyn GD, Marshall MK, Malloy R, Wiswell TE. Mainstream end-tidal carbon dioxide monitoring in the neonatal intensive care unit . Pediatrics 1998 ; 101 : 648–53
8.
Donn SM,Becker MA. Special ventilatory techniques and modalities I Patient Triggered ventilation. In Goldsmith JP, Karotkin EH (Eds) Assisted ventilation of the neonate. 4th Edition Saunders 2003, USA. pp 203 - 18
9.
Hagerty JJ, Kleinman ME, Zurakowski D, Lyons AC, Krauss B. Accuracy of a new low-flow SS capnography technology in newborns : A pilot study . J Perinatol 2002 ; 22 : 219–25
10.
McEvedy BA, McLeod ME, Kirpalani H, Volgyesi GA, Lerman J. End-tidal carbon dioxide measurements in critically ill neonates: a comparison of side-stream and mainstream capnometers. Can J Anaesth. 1990 ; 37(3):322-6
11.
Pascucci RC, Schena JA, Thompson JE. Comparison of a sidestream and mainstream capnometer in infants. Crit Care Med. 1989;17(6):560-2
12.
Hand IL, Shepard EK, Krauss AN, Auld PA. Discrepancies between transcutaneous and end-tidal carbon dioxide monitoring in the critically ill neonate with respiratory distress syndrome. Crit Care Med. 1989;17(6):556-9
13.
Moratin P, Lazarus G, Hartung E. [The reliability of proximal and distal gas samples in the capnometry of infants], Anaesthesist. 1992 ;41(6):307-12
14.
Lazzell VA, Burrows FA. Stability of the intraoperative arterial to end-tidal carbon dioxide partial pressure difference in children with congenital heart disease. Can J Anaesth. 1991;38(7):859-65.
15.
Nangia S, Saili A, Dutta AK. End tidal carbon dioxide monitoring – its reliability in neonates. Indian J Pediatr; 64: 389 -94
16.
From RP, Scamman FL. Ventilatory frequency influences accuracy of end-tidal CO2 measurements. Analysis of seven capnometers. Anesth Analg.1988; 67(9):884 -6
17.
Greer KJ, Bowen WA, Krauss AN. End-tidal CO2 as a function of tidal volume in mechanically ventilated infants. Am J Perinatol. 2003; 20(8):447-51
18.
Wu CH, Chou HC, Hsieh WS, Chen WK, Huang PY, Tsao PN. Good estimation of arterial carbon dioxide by end-tidal carbon dioxide monitoring in the neonatal intensive care unit. Pediatr Pulmonol. 2003; 35(4):292-5

JCDR is now Monthly and more widely Indexed .
  • Emerging Sources Citation Index (Web of Science, thomsonreuters)
  • Index Copernicus ICV 2017: 134.54
  • Academic Search Complete Database
  • Directory of Open Access Journals (DOAJ)
  • Embase
  • EBSCOhost
  • Google Scholar
  • HINARI Access to Research in Health Programme
  • Indian Science Abstracts (ISA)
  • Journal seek Database
  • Google
  • Popline (reproductive health literature)
  • www.omnimedicalsearch.com